Recall Z-0566-2013 ยท FPO Stretcher, Wheeled

Stryker Medical Division of Stryker Corporation

The Stryker Model 6390, Power-LOAD, is a power-loading cot fastener system designed to lift, lower, or steer ambulance cots into and out of the ambulance.

DateDec 19, 2012 (initiated Oct 11, 2012)
StatusTerminated
ReasonModel 6390 Power-LOAD systems built between September 12, 2012 and September 27, 2012 may have been built with a load switch slide that does not meet specifications causing the lifting ability of the arms to be negatively impacted.
Root causeComponent design/selection
ActionAll in process and Load Switch Slides within Stryker control were quarantined on 09/27/2012. Direct accounts were telephoned beginning 10/26/2012 with "Urgent Medical Device Correction" notices sent beginning on 11/30/2012. The notices explained the reason for the recall, hazards involved, and recommended actions. Stryker field service was to contact consignees to schedule replacement of the affected load switch slide. Consignees were asked to forward a copy of his notice to the new users and advise Stryker of their new location. Questions or concerns should be directed to (269) 389-6927 o
Quantity24
DistributionUSA, Nationwide including the states of PA, NM, MI, IA, CA, CO, WA, VA, IL, WI, NY, WY, MO, TX, and CT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113598POWER-LOAD COT FASTENER SYSTEM
Stryker Medical, Dec 20, 2011