Recall Z-0647-2013 ยท JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Siemens Healthcare Diagnostics

Immunoradiometric assay.

DateJan 7, 2013 (initiated Nov 28, 2012)
StatusTerminated
ReasonThe recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 (level G) lot 035. The affected calibrator vials can be identified by a darker yellow color than normal. The TSI3 (level A), TSI7 (level E), and TSIX (level H)
Root causeNonconforming Material/Component
ActionSiemens sent an Urgent Medical Device Recall letter dated November 2012 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product, complete and return the Urgent Medical Device Recall Effectiveness Check form included with the letter by fax to 302-631-7597. Customers were also instructed to determine their replacements needed, and to forward the recall letter to whomever they may have distributed the affected product. For questions customers were instructed to
QuantityLot 831=19 units (all foreign consignees) and Lot 832=2 units (all foreign consignees)
DistributionWorldwide Distribution - USA including CA, GA, IL, IN, MI, MN, NY, and Puerto Rico. Internationally to Agentina, Bolivia, Brazil, Canada, Chile, Guatemala, South Korea, Philippines, Paraguay, Taiwan, and Uraguay,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K872305IRMA-COUNT TSH RKTS1, RKTS2, RKTS5, RKTSX