Recall Z-0789-2013 ยท LXH Orthopedic Manual Surgical Instrument

Linvatec Corp. dba ConMed Linvatec

ConMed Linvatec Biomaterials, PopLok bone Punch PKL-35M, 3.5mm Orthopedic manual surgical instrument - Bone Punch

DateFeb 5, 2013 (initiated Nov 15, 2012)
StatusTerminated
ReasonConMed Linvatech recalled one lot of their PopLok bone Punch 3.5mm because the metal shaft diameter is 4.5mm when it should be 3.5mm.
Root causeEmployee error
ActionConMed Linvatech issued a "URGENT MEDICAL DEVICE RECALL NOTIFICATION" to its customers on November 15, 2012 for a recall that was conducted at the end user/distributor level. All recall notifications were sent FedEx next day delivery or equivalent method to notify all direct consignees. The notifications identified the affected product along with reason for the recall health risk. All accounts were asked to immediately check their facility for the affected product, stop using any of the product with specified lot number, and to segregate the product. The letter also asked customers to return
Quantity8
DistributionWorldwide Distribution - USA including California and Internationally to Spain and Korea