Recall Z-0641-2013 ยท JJW Urinalysis Controls (Assayed And Unassayed)

Siemens Healthcare Diagnostics Inc

Siemens Chek-Stix Combo Pak (Distributed OUS) Catalog Number: 1364E Chek-Stix Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes

DateJan 4, 2013 (initiated Nov 21, 2012)
StatusTerminated
ReasonWhen these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F).
Root causeNonconforming Material/Component
ActionSiemens issued An Urgent Field Safety Notice via e-mail to all affected SiemensHealthcare Diagnostics Regional offices on November 21, 2012 both in the United States and foreign accounts. This notice informed customers of the issue with the CLINITEK Atlas and Chek-Stix Positive Urinalysis Control Solutions When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F). You can use the reconstituted positive control solution for up to eight hours if the solution is held at
Quantity4642
DistributionWorldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K943780NEGATIVE CONTROL STRIPS FOR URINALYSIS (CHEX-STIX, CLINITEK AUTO 2000, CLINITEX ATLAS)
Heraeus Kulzer, Inc., Oct 31, 1994