Recall Z-0752-2013 ยท HWC Screw, Fixation, Bone

Extremity Medical LLC

Extremity Medical, IO FiX System Accessory. Catalog Number 101-00010. This screw measuring device is an accessory to the IO FiX Screw and Washer System that is used to measure the length of the screw by placing the depth gauge over the guidewire and down to the bone. Orthopedic device.

DateJan 30, 2013 (initiated Mar 22, 2012)
StatusTerminated
ReasonLag screw measurement is short (approximately 5 mm) when using current screw measuring device (p/n 101-00010) per the IO FIX Surgical Technique Rev C. which can lead to possible inappropriate implant size selection (e.g. shorter screw size selected).
Root causeDevice Design
ActionExtremity Medical LLC sent e-mail notification dated March 22, 2012, to all affected customers. The email identified the product the problem and the action to be taken by the customers. The email includes a buleeting which informes the customer that an upgrade to the existing instrumentation is being released. The upgrade will replace the existing gauge. Updated pdf versions of the technique guide are also included. Customers are instructed to return the recalled device after swapping out the recalled device with the replacement unit. Customers will be contacted by Extremity and Sales Reps
Quantity93 units
DistributionWorldwide Distribution -- Nationwide Distribution including the states of AR, AZ, CA, FL, GA, IL, IN, MD, MT, NC, NE, NH, NJ, NY, OH, PA, TN, TX, WI and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101700IO FIX HAND, IO FIX FOOT, EXTREMITY MEDICAL SCREW SYSTEM
Extremity Medical, LLC, Oct 12, 2010