Recall Z-0625-2013 ยท FRN Pump, Infusion

Hospira Inc.

Plum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 12348 - software version 10.3; b) list 12618 - software version 11.3; c) list 20678 - Hospira MedNet software; Product Usage: The Plum A+3 is a dual-line triple channel volumetric infusion system that provides precise delivery of multiple i.v. therapies acro

DateDec 31, 2012 (initiated Dec 7, 2012)
StatusTerminated
ReasonThe door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.
Root causeDevice Design
ActionHospira Inc. sent "URGENT DEVICE FIELD CORRECTION" letters, dated December 7, 2012, to all affected customers. The letters identified the products, problem and action(s) to be taken until doors have been replaced. The customers were instructed to read the information provided carefully, including instructions to inspect the door roller assembly prior to loading a cassette for signs of damage. Customers are to complete the attached Reply Form and return it to the firm via fax at 1-877-650-8359. Customers were instructed to notify their consignees if they have further distributed the affected
Quantity28,599 units
DistributionWorldwide Distribution -- USA, including Puerto Rico, and the countries of Austria, Australia, Argentina, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, Finland, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021350ABBOTT PLUM A+3 MULTICHANNEL INFUSION PUMP, MODEL 12348
Abbott Laboratories, May 14, 2002
K042081PLUM A+ INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE , AND PLUM A+3 INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE
Hospira, Inc., Aug 24, 2004
K052052HOSPIRA PLUM A+ INFUSION PUMP SYSTEM V11.5, HOSPIRA PLUM A+ 3 INFUSION PUMP SYSTEM, V11.5
Hospira, Inc., Aug 24, 2005