Recall Z-0748-2013 ยท BTL Ventilator, Emergency, Powered (Resuscitator)

GE Healthcare, LLC

GE Healthcare - Giraffe and Panda T-Piece Resuscitation System The T-piece Resuscitation System [ventilator, emergency powered resuscitator]/Panda Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants.

DateFeb 8, 2013 (initiated Oct 15, 2012)
StatusTerminated
ReasonOxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
Root causeEmployee error
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated October 15, 2012, with a return receipt to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer along with the product correction. Required Action for Users: It is recommended that a qualified biomedical technician or respiratory therapist identify the affected resuscitation system and locate the air and oxygen fittings and labels at the back of the unit. With the table provided, identify the appropriate fitting for their country and location and verify the air fi
Quantity276 units total, 109 of those nationwide.
DistributionWorldwide Distribution - Worldwide Distribution - USA including AL, AZ, CA, CO, FL,GA, IA, ID, IL, KY, MA, MD, MI, MN, MO, MS, MT, NC, ND ,NE, NM, NV, NY, OH, OR, PA ,TX, VT, WA, WI and WV; and the countries of Australia, Belgium, Canada, China, France, New Zealand, Poland, Saudi Arabia, Switzerla
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070210GIRAFFE AND PANDA T- PIECE RESUSCITATION SYSTEM
Ohmeda Medical, Apr 20, 2007