Hospira Inc.
Plum A+ Single Channel Infusion Pumps with Hospira Mednet Software and a Connectivity Engine peripheral module that provides wired Ethernet and wireless local area networking, allowing the MedNet software to download drug libraries to the infuser and enable the auto-programming feature; Hospira, Inc., Lake Forest, IL 60045; pump list #20679 and 20792, module list number 20677. The Plum A+ Infusio
| Date | Dec 7, 2012 (initiated Oct 26, 2012) |
| Status | Terminated |
| Reason | The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knob is turned clockwise. |
| Root cause | Device Design |
| Action | The firm, Hospira Inc. sent "URGENT DEVICE FIELD CORRECTION" letters dated October 26, 2012 to all affected customers. The letters identified the products, problem and action(s) to be taken. The customers were instructed to read the information provided carefully, complete the attached Reply Form and return it to the firm via fax at 1-877-650-8667. The letter states that the System Operating Manual will be updated to clarify the instructions for setting the volume, and that the updated manuals will be sent as soon as they are available. Customers were instructed to notify their consignees if t |
| Distribution | Worldwide Distribution - USA (nationwide) including Puerto Rico, and the countries of Australia, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K042081 | PLUM A+ INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE , AND PLUM A+3 INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE Hospira, Inc., Aug 24, 2004 |