ClearanceBrief
Recall Z-0576-2013 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

DePuy Orthopaedics, Inc.

SOL SYS 8IN CALC 1.5/16.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SO

DateDec 20, 2012 (initiated Dec 3, 2012)
StatusTerminated
ReasonAn Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a
Root causeDevice Design
ActionDeputh Orthopaedics sent an Urgent Information - Medical Device Recall Notice dated December , 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The notification letter instructed consignees to please cease further distribution or use of products immediately. The purpose of the communication was to inform them of this recall and request acknowledgement of receipt of this letter by their facility. Reconciliation forms should be returned to your DePuy Sales Representative or faxed to the following Fax Number: 574-371-49
Quantity1732 total all product codes
DistributionNationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060581DEPUY SOLUTION SYSTEM HIP PROSTHESIS
DePuy Orthopaedics, Inc., May 19, 2006