ClearanceBrief
Recall Z-0496-2013 ยท LKK Pump, Infusion, Implanted, Programmable

Medtronic Neuromodulation

SynchroMed EL, models 8626-10, 8626L-10, 8626-18, 8626L-18, 8627-10, 8627L-10, 8627-18, 8627L-18. implantable infusion pump. The pump is part of an implanted infusion system that consists of a SynchroMed EL pump and a catheter. The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The SynchroMed EL Infusion System with a 10

DateDec 13, 2012 (initiated Nov 9, 2012)
StatusTerminated
ReasonMedtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps.
Root causeNonconforming Material/Component
ActionThe firm Medtronic, sent an "Urgent medical Device Safety Notification" dated November 2012, to its customers beginning November 09, 2012. The letter described the product, problem and actions to be taken. The customers were instructed to follow the recommendations provided in the letter and complete and return the PHYSICIAN REPLY FORM via mail at Medtronic Neuromodulation Quality Assurance, 7000 Central Ave NE, Mailstop RCC150, Minneapolis, MN 55432; fax to: 800-897-3899 or email to: neuro.quality@medtronic.com. If you have questions, please contact Medtronic Neuromodulation Technical Servi
Quantity174,700 devices total
DistributionWorldwide distribution: USA (nationwide) and countries including: Canada and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.