Recall Z-0495-2013 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips Healthcare Inc.

Philips IntelliVue Information Center/Client and Philips IntelliVue Information Center iX The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Sof

DateDec 10, 2012 (initiated Nov 9, 2012)
StatusTerminated
ReasonInternal audio cable may have diminished retention force due to damaged contacts, which could lead to intermittent or loss of audible alarms.
Root causeComponent design/selection
ActionPhilips sent an URGENT - Medical Device Correction letter dated November 9, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Philips is conducting a voluntary correction to replace the audio cable on affected devices. Customer were instructed to follow the "Action to be taken by CustomerUser" section of the instructions. Customers with wquestions were instructed to contact their Phililps representataive or contact the Customer Care Solutions Center at (800) 722-9377. For questions regarding this recall call 978-68
Quantity3165
DistributionWorldwide Distribution - USA (nationwide) and Canada.