Recall Z-0840-2013 ยท JOW Sleeve, Limb, Compressible

Stryker Sustainability Solutions

STRYKER Sustainability Solutions, AC Power Adaptor for Restep Deep Vein Thrombosis (DVT) System, an external compression system. PUMPKIT (Pump Kit), and RSP159-01 (AC Power Adaptor).

DateFeb 20, 2013 (initiated Aug 30, 2012)
StatusTerminated
ReasonReports that the AC Power Adaptor may not operate due to wires coming loose in the unit.
Root causeProcess control
ActionStryker Sustainability Solutions sent a Urgent Medical Device Removal Notification letter dated September 5, 2012, to all affected customers. The letter identified the product, problem and the action to be taken by the customer. The firm notified that affected AC Power Adaptors would be replaced with improved Adaptors. Any affected lots would be removed and shipped back to Stryker Sustainability Solutions for final disposition. Customers were instructed to fill out the attached Recall Effectiveness Form even if no affected product was found and fax to Stryker Sustainability Solutions at
Quantity1,311
DistributionDistributed in Arizona and New Jersey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090308RESTEP DVY SYSTEM, MODEL RSP-101
Stortford Medical, LLC, May 1, 2009