Recall Z-0594-2013 ยท MAX Intervertebral Fusion Device With Bone Graft, Lumbar

Zimmer, Inc.

Zimmer Spine Ardis Inserter, Rx, non-sterile

DateDec 22, 2012 (initiated Nov 12, 2012)
StatusTerminated
ReasonZimmer Spine has received reports of the PEEK Ardis Interbody Spacer breaking into fragments when too much lateral force is applied to the Ardis Inserter. If the PEEK Ardis Interbody Spacer breaks during implantation, surgical delays of up to 60 minutes could occur. Sharp edges of the broken implant may contact and damage adjacent structures, requiring surgical intervention. Observed health risks
Root causeDevice Design
ActionZimmer Spine, sent a "MEDICAL DEVICE CORRECTION" memo dated November 12, 2012 to all consignees / customers describing product, problem and actions to be taken including revised surgical technique. Beginning the week of December 17, 2012, Zimmer began the immediate retrieval of the affected units as a Class I recall. December 20, 2012 nation wide press outlining the recall was issued. Customers may report adverse events or quality problems experienced with the use of these products to Zimmer Spine by calling 866-774-6368 between the hours of 8AM and 5PM CST or by email at qualitycompliance@z
Quantity315 (172 in USA)
DistributionWorldwide distribution: USA (nationwide) including states of: AR, AZ, CA, CO, CT, DC, FL, GA, ID, LA, MD, MI, MN, MO, NC, NV, NY, OH, OK, PA, TN, TX, WI, and WV; and countries of: Australia, El Salvador, France and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073202ARDIS SPACER
Abbott Spine, Inc., Jan 30, 2008