Recall Z-0440-2013 ยท LKF Cannula, Manipulator/Injector, Uterine

Cooper Surgical, Inc.

Cooper Surgical Advincula Arch , Reusable Uterine Manipulator Handle Ref: UMH750 The intended use of the RUMI Arch is in an operative endoscopy (laparoscopy) where a uterus is present and where positioning of the uterus, fallopian tubes and ovaries is desirable. These types of surgeries include laparoscopic tubal ligation, and/or operative laparoscopy

DateNov 26, 2012 (initiated Oct 31, 2012)
StatusTerminated
ReasonExcessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle
Root causeDevice Design
ActionCooperSurgical issued notification on October 31, 2012, to all affected customers Fed X . The notification identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return units for replacement. Customers with questions should contact Customer Serivce at 1-203-601-9818 or 800-243-2974. Customers were asked to fill out the attached Replacement of Recalled Product Form and fax it to 1-800-262-0105. For questions regarding this recall call 203-601-9818.
Quantity1215 total
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Australia, Belgium, Canada, Czech Republic, Germany, Italy, Spain, France, United Kingdom, Japan, New Zealand, South Korea, Saudi Arabia, Sweden, Singapore, Slovenia, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K932115ROWDEN UTERINE MANIPULATOR INJECTOR AND RUMI
Blairden Precision Instruments, Inc., Jun 24, 1994