ClearanceBrief
Recall Z-0709-2013 ยท NKB Thoracolumbosacral Pedicle Screw System

DePuy Spine, Inc.

DePuy Spine VIPER POLY SCREW 5.5 9 X 80MM Tl Product Code: 1867-15-980 Spinal Implant Component.

DateJan 24, 2013 (initiated Nov 7, 2012)
StatusTerminated
ReasonPolyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head
Root causeComponent change control
ActionDePuy Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated November 7, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, quarantine the material and return remaining inventory. Customers were advised to complete the enclosed Business Reply Form and fax back to 508-828-3762 or email to lregan3@its.jnj.com. For questions call Linda Regan @508-977-6606.
Quantity34 uits
DistributionWorldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101993VIPER SYSTEM, EXPEDIUM SYSTEM
Depuy Spine, Inc., Dec 7, 2010