ClearanceBrief
Recall Z-0764-2014 ยท HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)

The Anspach Effort, Inc.

MCA7-15SB - 1.5MM Fluted Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.

DateJan 15, 2014 (initiated Mar 29, 2012)
StatusTerminated
ReasonThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
Root causeLabeling Change Control
ActionFirm is performing a Medical Device Recall by sending customers a recall letter and requesting the return of Anspach Micro Curved Attachments and Dissection Tools currently on hand.
Quantity6439 total
DistributionWorldwide Distribution: US (nationwide) and Internationally to: AU, BE, CA, CH, ES, FR, GB, IT, JP, NL, SE, and ZA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042783ANSPACH MINIMAL ACCESS SPINAL ATTACHMENT (MASA) SYSTEM
The Anspach Effort, Inc., Dec 17, 2004