Recall Z-0814-2013 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
GE Healthcare, LLC
GE Healthcare, Dash Port 2 Docking Station. Product Usage: is a docking station (quick mount/dismount base) for a Dash patient monitor. The system is to monitor physiologic parameter data on adult, pediatric, and neonatal patients.
| Date | Feb 14, 2013 (initiated Jun 13, 2012) |
| Status | Terminated |
| Reason | GE has become aware of multiple issues affecting the Dash 3000/4000/5000 and DashPort 2 monitor. This report of correction encompasses the DashPort 2 device, in which the DashPort 2 stops communicating with the Dash, resulting in Loss of Monitoring at the CIC. The DashPort 2 status LEDs in the back of the unit indicate Unity Network and DashPort 2 to Dash communication has stopped. |
| Root cause | Nonconforming Material/Component |
| Action | GE Healthcare sent an Important Product Information Letter dated June 13, 2012 via FedEx to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter indicated that the motherboard would be replaced, on request, should the customer continue to experience the issue. A second letter dated August 30th 2012 was issued informing consignees that the correction that was initially proposed did not fully correct the issues. Therefore, the motherboard replacement field action had been suspended. It further explained that the correction will be prov |
| Quantity | 10,497 |
| Distribution | Worldwide Distribution and US (nationwide) including DC and the countries of ALGERIA, ARAB EMIRATES, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BRAZIL, CANADA, CHINA, CYPRUS, CZECH REPUBLIC, DENMARK, EGYPT, FRANCE, GERMANY, INDIA, ISRAEL, ITA LY, JAPAN, JORDAN, KUWAIT, MOROCCO, NETHERLANDS, NEW ZEALAN |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K073462 | MODIFICATION TO DASH 3000, 4000, 5000 MONITOR AND ACCESSORIES Ge Medical Systems Information Technologies, Jan 11, 2008 |