Recall Z-0411-2013 ยท ETD Tube, Tympanostomy

Gyrus Acmi, Incorporated

PARPARELLA-TYPE VENT TUBE, 1.02 MM, SILICONE, REF 240044, QTY 6, Rx Only, STERILE EO, GYRUS ACMI INC., Tympanostomy tube

DateNov 20, 2012 (initiated Oct 17, 2012)
StatusTerminated
ReasonOne lot of tympanostomy tubes were manufactured with silicon that was found to be contaminated with ethylene vinyl acetate copolymer.
Root causeMaterial/Component Contamination
ActionThe firm, Olympus, sent an "URGENT: Medical Device Recall" letter dated October 15, 2012 to their consignees/customers. The letter describes the product, problem and actions to be taken. The customers were instructed to inspect their stock to confirm if any of the affected lot of product remains in their possession; immediately cease any further use of any affected product; remove it from their stockroom and quarantine it until it is shipped back to Gyrus ACMI, Inc.; call their Gyrus ACMI customer service representative (800-773-4301) to obtain a Returned Goods Authorization and promptly retu
Quantity10 boxes/60 units
DistributionWorldwide distribution: USA state of Kansas and country of Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.