ClearanceBrief
Recall Z-1071-2013 ยท FMQ Restraint, Protective

J T Posey Company

Posey Biothane Restraint Cuffs, Catalog Number 2900. The device is intended to limit a patient's movement to the extent necessary for treatment or protection of the patient or others.

DateApr 9, 2013 (initiated Oct 31, 2012)
StatusTerminated
ReasonThe firm voluntarily recalled Biothane Restraint Cuffs, Catalog No. 2900, that did not meet the required specifications for tensile strength. Failure of these rivets could allow the cuff to become disconnected from the connecting strap and allow patients to regain free use of the limb.
Root causeProcess design
ActionJ. T. Posey sent an Medical Device Recall (Removal) letter to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter instructed the customer that Posey Company is voluntarily recalling all units of Posey BiothaneQD Cuffs, Catalog Number 2900. All lots of this product are being recalled. Reason for recall was listed:"Posey has identified rivets on the Biothane~ Restraint Cuffs, Catalog No. 2900, that did not meet the required specifications for tensile strength. Failure of these rivets could allow the cuff to becom
Quantity1673 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.