Recall Z-0463-2013 ยท GGP Test, Qualitative And Quantitative Factor Deficiency

Sekisui Diagnostics Llc

Sekisui Diagnostics, Spectrolyse PAI-1 In-Vitro Diagnostic Use for the quantitative determination of Plasminogen Activator Inhibitor Type 1 in human plasma. Catalog # 101201. Spectrolyse PAI-1 is intended for the quantitative determination of Plasminogen Activator Inhibitor Type-1 (PAI-1) activity in human plasma. The test is for in vitro diagnostic use.

DateDec 3, 2012 (initiated Oct 24, 2012)
StatusTerminated
ReasonSpectrolyse PAI-1 activity assay, producing lower than expected absorbance values 0 IU/ml standard reducing the slope of the standard curve requiring more frequent sample dilutions.
Root causeNonconforming Material/Component
ActionSekisui Diagnostics issued an "URGENT PRODUCT RECALL NOTICE" dated October 24, 2012 to all consignees. The letter instructed consignees to immediately stop using the specified lot of product, discard any remaining inventory and return the enclosed response form. Contact the Customer Service Department at 800-637-3375, ext. 10, 12, or 31 for assistance with this recall.
Quantity63 kits
DistributionNationwide Distribution-including the states of FL, IN, NC, OH, UT, and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063323SPECTROLYSE PAI-1, MODEL 101201
American Diagnostica, Inc., Aug 15, 2007