Recall Z-0471-2013 ยท HTO Reamer
Biomet, Inc.
OSS Reamer Sleeve REF132-472661 KNEE REAMER SLEEVE 8.0 THRU 12.5 STAINLESS STEEL The OSS Knee Reamer Sleeves are primarily used to guide the bullet tip of cylindrical reamers during the tibial preparation in OSS procedures. They are reusable.
| Date | Dec 4, 2012 (initiated Nov 1, 2012) |
| Status | Terminated |
| Reason | Biomet has initiated this action following an investigation which identified that the thin shaft reamer may get stuck and not slide into the sleeve. |
| Root cause | Nonconforming Material/Component |
| Action | The firm, Biomet, sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter dated November 1, 2012 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately locate, discontinue use of and remove the product from circulation; and return recalled devices with directions to distribute notices, if product was further distributed; and to complete and return the enclosed FAX Back Response Form within three (3) business days to FAX # 574-372-1683. Distributors were charged with locating and returning products to Biomet ATTN: Return |
| Quantity | 46 |
| Distribution | Worldwide distribution: USA (nationwide) and countries of: Japan and Netherlands. |
| FDA record | Open on fda.gov |