Recall Z-0628-2013 ยท FNL Bed, Ac-Powered Adjustable Hospital
Stryker Medical Division of Stryker Corporation
Medical Surgical Bed, Model 3002SEX Medical Surgical Bed The S3" MedSurg Bed, is intended to support and transport patients within the Med/Surg and Critical Care hospital environments. The bed is typically used in pre-op, post-op and recovery areas of hospital facilities. The intended user for this product is both Health Care Providers (HCPs: nurses, nurses aides, and medical doctors) and human p
| Date | Jan 3, 2013 (initiated Dec 18, 2012) |
| Status | Terminated |
| Reason | Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. The affected beds may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes. |
| Root cause | Process control |
| Action | Stryker sent an Urgent Medical Device Correction letterdated December 18, 2012, to all affected customer. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate the affected product and ensure that the brakes are holding properly. Consignees were directed to contact Stryker (1-800-327-0770, option 2) to make arrangements to replace all brake rings and continue to routinely inspect brake functionality. Customers were asked to Fax (269-488-8691) the enclosed acknowledgment and Stryker field service will contact custo |
| Quantity | 9 |
| Distribution | Worldwide Distribution - Nationwide and Canada |
| FDA record | Open on fda.gov |