Recall Z-0200-2013 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

GE Healthcare, LLC

GE Healthcare Brivo XR 285 Mobile general purpose radiographic imaging of the human head and body.

DateNov 16, 2012 (initiated Sep 20, 2012)
StatusTerminated
ReasonIt was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54.
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare will remedy the defect or bring the product into compliance with each applicable Federal and IEC standard. A Service Representative will update the firmware on the system to address the issue. Field Modification Instructions (FMI) 10870, 10872, and 10873 are describing the rework plan the GEHC Engineers will make to the imaging systems compliant with the applicable performance standard. Further questions please call (262) 513-4122.
Quantity364 total units installed in US
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.