Recall Z-0659-2013 ยท MAX Intervertebral Fusion Device With Bone Graft, Lumbar

Zimmer, Inc.

The TM Ardis Interbody System Inserter Stainless Steel, Rubber, Silicone Non Sterile Rx only. The TM Ardis Inserter is intended for delivery of the TM Ardis Implant into the cleared disc space. The implant is secured to the inserter via finger-tightening the knob of the inserter shaft allowing the lateral grasping arms to secure the implant. Once the implant is inserted and its position confirmed

DateJan 9, 2013 (initiated Sep 20, 2012)
StatusTerminated
ReasonZimmer Spine is has received complaints of difficulty in turning the implant release knob of the TM Ardis Inserter after final positioning of the TM Ardis interbody spacer implant. There has been two reports where the inserter instrument was inadvertently removed from patient with the implant still engaged. The correction includes revised directions explaining the technique to remove the implant
Root causeDevice Design
ActionZimmer sent a "MEDICAL DEVICE CORRECTION" letter dated September 26, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Risk Manager Confirmation Form was attached for customers to complete and return. Questions and concerns should be addressed to Zimmer Spine Customer Service at 866-774-6368. When re-design is approved, Zimmer will remove and replace all affected devices.
Quantity170
DistributionUSA nationwide; Europe, Sweden
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113561TM ARDIS INTERBODY SYSTEM TM ARDIS INTERBODY SYSTEM INSTRUMENATION
Zimmer Trabecular Metal Technology, May 29, 2012