Recall Z-0466-2013 ยท JAK System, X-Ray, Tomography, Computed

Philips And Neusoft Medical Systems Co., Ltd.

NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

DateDec 4, 2012 (initiated Oct 18, 2012)
StatusTerminated
ReasonThe four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
Root causeDevice Design
ActionPhilips and Neusoft Medical Systems Co., Ltd. sent an Urgent Device Correction NeuViz Dual, NeuViz 16 letter, dated Oct. 18, 2012, to all affected US customers and distributors. The letter described the product, problem, and actions to be taken by the customers. Customers are advised to: 1. Have a mechanic check the patient table top screws fixing. Detailed instructions are attached as Annex with the letter. 2. If it is found that table top screws are improperly assembled, stop using the product and contact local Service personnel for repair. Philips and Neusoft Medical Systems Co., Ltd
Quantity20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series
DistributionNationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062451NEUVIZ DUAL MULTI-SLICE CT SCANNER SYSTEM
Neusoft Medical Systems Co., Ltd., Sep 28, 2006
K071308MODIFICATION TO NEUVIZ DUAL MULTI-SLICE CT SCANNER SYSTEM
Neusoft Medical Systems Co., Ltd., Aug 16, 2007
K090173NEUVIZ 16 MULTI-SLICE CT SCANNER SYSTEM
Neusoft Medical Systems Co., Ltd., Feb 5, 2009
K092742NEUVIZ 16 MULTI-SLICE CT SCANNER SYSTEM
Neusoft Medical Systems Co., Ltd., Sep 22, 2009