Recall Z-0259-2013 ยท FTL Mesh, Surgical, Polymeric

American Medical Systems, Inc.

AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS Male TO Sling System (Intl). The AMS AdVance Male Sling System is intended for the placement of a suburethral sling for the treatment of male stress urinary incontinence (SUI).

DateNov 7, 2012 (initiated Oct 19, 2012)
StatusTerminated
ReasonAmerican Medical Systems (AMS) is issuing a voluntary recall on a number of lots of the AdVance Male Sling System (AdVance). AMS has recently received a small number of complaints regarding the resorbable suture in the AdVance sling not lasting for the entire 15 month shelf life of the product. An internal investigation has been conducted and confirmed that some sutures degraded prior to the
Root causeDevice Design
ActionAmerican Medical Systems sent an "URGENT MEDICAL DEVICE RECALL" dated October 18, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter instructs customers to return any existing stock of the recall product and to complete and return the Product Recall Acknowledgment Form. Contact AMS Customer Service at 1-800-328-3881 for additional information or questions. .
Quantity4872
DistributionWorldwide Distribution-USA (nationwide) including DC and Puerto Rico and the countries of Austria, Australia, Brazil, Canada, Germany, Estonia, Spain, Finland, France, Great Britain, Hong Kong, Croatia, India, Morocco, Mexico, Malaysia, Netherlands, New Zealand, Panama, Poland, Russian Federation, S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053371AMS MALE TRANSOBTURATOR SLING SYSTEM
American Medical Systems, Inc., Feb 3, 2006