Recall Z-0407-2013 ยท KAK Probe, Ent

Vital Signs Colorado Inc.

Vital Signs Disposable General Purpose 9 French Temperature Probe For monitoring of core temperature

DateNov 20, 2012 (initiated Aug 17, 2012)
StatusTerminated
ReasonThere is a potential for epistaxis during use of the Disposable General Purpose 9 French Temperature Probe.
Root causeDevice Design
ActionThe firm,GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter and Confirmation form on August 17, 2012 via Fed Ex to its consignees/customers. The letter describes the product, problem and action to be taken. The customers were instructed to do the following: Do not use the Disposable General Purpose 9 French Temperature Probe P/N 1016 or 1016EU; return all unopened product(s) to Vital Signs Devices (VSD) 11039 East Lansing Circle, Englewood, CO 80112-note: replacements and/or refunds discussed in letter are available only in exchange for the return of any unopened product(s); comp
Quantity243,685 units
DistributionWorldwide distribution: USA (nationwide) including states of: AR, CA, FL,GA, IL, LA, MA, MI, MO, NC, NH, NY, OH, OK, PA, RI, SE,TX, and WA; and countries of: AUSTRALIA, BARCELONA, GERMANY, NEW ZEALAND, NETHERLANDS, SWITZERLAND, and UNITED KINGDOM.
FDA recordOpen on fda.gov