Recall Z-0465-2013 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System

DateDec 3, 2012 (initiated Jun 20, 2012)
StatusTerminated
ReasonFirm has become aware of an unintended behavior when using syngo.plaza and issued an Update Instruction. The Update Instruction SY050/12/S provides a software update and addresses the following issues: - Patient Rename messages for the same patient are being continually processed for several days. This fills up the 'pcvdbsrv' log and potentially affects performance of the system. - Data could
Root causeSoftware design
ActionThe firm, Siemens, sent "Update Instruction SY050/12/S" on June 20, 2012 to its consignees/customers. The Update Instructions described the product, problem and actions to be taken. It also provides a software update. If you have any questions call 610-219-4834 or email: anastasia.mason@siemens.com.
Quantity15
DistributionNationwide distribution: USA including states of: CA, FL, GA, MA, MN, MO, NJ, NY, OH, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093612SYNGO.PLAZA
Siemens AG Medical Solutions, Mar 4, 2010