Recall Z-0390-2013 ยท HWT Template

Zimmer, Inc.

00789502601 VerSys Femoral Head Provisional 26mm -3.5 00789502602 VerSys Femoral Head Provisional 26mm +0 00789502603 VerSys Femoral Head Provisional 26mm +3.5 00789502604 VerSys Femoral Head Provisional 26mm +7 00789502605 VerSys Femoral Head Provisional 26mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stabilit

DateNov 19, 2012 (initiated Oct 10, 2010)
StatusTerminated
ReasonSurgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat
Root causeDevice Design
ActionThe firm, Zimmer, sent notification by email to consignees beginning October 10, 2012 with a follow-up letter titled "URGENT MEDICAL DEVICE CORRECTION" dated October 16, 2012 to its consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to complete a spreadsheet of current users and contact information and immediately provide this information via email to: CorporateQuality.Postmarket@zimmer.com, and review and become familiar with the content and recommendations of the attached letter sent to the customers. For questions
Quantity128,946 all lots, all sizes
DistributionWorldwide distribution: USA (nationwide) and countries of: Asia, Barbados, Canada, Europe, Mexico, Puerto Rico, San Juan, and South America.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.