Recall Z-0482-2013 ยท FRN Pump, Infusion

Hospira Inc.

Plum A+ Hyperbaric Single Channel Infusion Pumps with Hospira MedNet Software and a Connectivity Engine peripheral module that provides wired Ethernet and wireless local networking, allowing the MedNet software to download drug libraries to the infuser and enable the auto-programming feature; Hospira, Inc., Lake Forest, IL 60045; pump list number 11005, module list number 11006 Product Usage: T

DateDec 7, 2012 (initiated Oct 26, 2012)
StatusTerminated
ReasonThe volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knob is turned clockwise.
Root causeDevice Design
ActionThe firm, Hospira Inc. sent "URGENT DEVICE FIELD CORRECTION" letters dated October 26, 2012 to all affected customers. The letters identified the products, problem and action(s) to be taken. The customers were instructed to read the information provided carefully, complete the attached Reply Form and return it to the firm via fax at 1-877-650-8667. The letter states that the System Operating Manual will be updated to clarify the instructions for setting the volume, and that the updated manuals will be sent as soon as they are available. Customers were instructed to notify their consignees if t
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico, and the countries of Australia, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081412PLUM A+ HYPERBARIC INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE
Hospira, Inc., Dec 17, 2008