Recall Z-0347-2013 ยท CEM Electrode, Ion Specific, Potassium

Beckman Coulter Inc.

UniCel DxC 880i, 860i, 680i, and 660i Integrated Systems, Part Numbers: A64871, A64903, A64935, A59102. The UniCel DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The UniCel DxC series of instruments is a DxI Immunoassay System integrated with a chemistry

DateNov 19, 2012 (initiated Aug 1, 2011)
StatusTerminated
ReasonThe recall was initiated because Beckman Coulter has confirmed that the UniCel Dxl and DxC Access Immunoassay Systems had field modifications to the aspirate and dispense probe assemblies.
Root causeProcess design
ActionBeckman Coulter sent a field modification was released in August 2011 which included: - Installing a new aspirate and dispense tubing routing bracket. - Replacing existing aspirate and dispense probe tubing with modified tubing to help eliminatemisrouting of aspirate and dispense probe tubing. During installation the FSE informed the customer: -That aspirate and dispense tubing are different lengths. - That probes are not interchangeable with other instruments once tubing has been modified. - To route probe tubing by referring to the labels on the tubing. An updated service document, Routine
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Canada, AE, AU, BG, BW, CH, CN, CZ, DE, ES, FR, GB, HK, HR, HU, IN, IT, JO, JP, LB, MA, MX, NL, PL, QA, RO, RU, SK, TN, TR, ZA, AO, AR, AT, BE, CO, DZ, FI, IE, IL, KR, KW, LI, NA, NO, NZ, PF, PH, PR, SE, SG, SZ, TH, TW, VE, VN, and YT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042291UNICEL DXC 600 AND 800 SYNCHRON SYSTEMS
Beckman Coulter, Inc., Nov 12, 2004