Recall Z-0381-2013 ยท LXH Orthopedic Manual Surgical Instrument

Stryker Spine

Stryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct. Allendale NJ 07401-1677 USA +1-201-780-8000 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks

DateNov 19, 2012 (initiated Aug 9, 2012)
StatusTerminated
ReasonMultiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.
Root causeNonconforming Material/Component
ActionStryker Spine sent an Urgent Medical Device Removal Notification letter/Product Accountability Forms dated August 9, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. For questions regarding this recall call 201-760-8150.
Quantity19 units
DistributionWorldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.