Recall Z-0157-2013 ยท DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass

Sorin Group USA, Inc.

Gish 3/8 x 3/8" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15. Product Usage: The connectors are indicated for use in connecting flexible tubing used in extracorporeal blood or fluid circuits. The connectors are distributed as sterile stand alones and bulk non-sterile.

DateOct 25, 2012 (initiated Sep 14, 2012)
StatusTerminated
ReasonDue to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.
Root causeProcess control
ActionSorin Group USA, Inc. sent a Product Recall letter to all affected customers by certified mail on September 14, 2012. The letter identified the affected products, problem and actions to be taken. Customers were instructed to discontinue use of the affected product, return all unused product to the firm and to complete and return a Response Form via fax or email. For further questions, contact the firm at 1-800-650-2623.
Quantity1,640 units
DistributionUS Nationwide Distribution in the states of: AL, AR, CA, CO, CT, FL, IL, MI, MN, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K833322TUBING CONNECTORS EC-2XXX SERIES
Gish Biomedical, Inc., Dec 16, 1983