ClearanceBrief
Recall Z-0387-2013 ยท HWC Screw, Fixation, Bone

Stryker Howmedica Osteonics Corp.

Stryker Orthopaedics Torx 6.5mm Cancellous Bone Screw Ref 2030-6525-1 The 6.5mm Cancellous Bone Screws are used in THA when the surgeon elects to use a cluster/multi-hole shell in the acetabulum. The screws are intended to provide initial cup fixation and help prevent spin-out of the cup due to physical loading from the patient. They are intended as supplemental fixation.

DateNov 19, 2012 (initiated Aug 17, 2012)
StatusTerminated
ReasonStryker has received reports from the field indicating that certain lots of 6.5mm Cancellous Bone Screws are either associated with a product mix or potentially associated with a product mix. Packaging associated with a 35 mm screw may contain a 25mm screw. Conversely, packaging associated with a 25mm screw may contain a 35mm screw.
Root causeMixed-up of materials/components
ActionThe firm, Stryker Orthopaedics, notified Domestic locations of this action by e-mail on August 6, 2012 and foreign locations of this action by e-mail on August 7, 2012. Stryker follow-up by sending an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" dated August 17, 2012 to their consignees/customers via Fed Ex with return receipt. The letter described the product, problem and actions to be taken. The customers were instructed to quarantine and return affected product to Attn: Regulatory Compliance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah, New Jersey, 07430 (attach fluorescent orange sti
QuantityApproximately 72 units
DistributionWorldwide distribution: USA (nationwide) and countries including: Canada, Belgium, Finland, Germany, Greece, Hong Kong, Ireland, Italy, Japan, Poland, Portugal, Spain, Turkey and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K894124OMNIFIT SELF TAPPING CANCELLOUS BONE SCREWS
Osteonics Corp., Sep 11, 1989