Recall Z-0293-2013 ยท LMF Agent, Absorbable Hemostatic, Collagen Based
Ethicon, Inc.
ETHICON SURGIFLO HEMOSTATIC MATRIX Distributed by ETHICON ETHICON, Inc. Somerville, NJ 08876, USA SURGIFLO Hemostatic Matrix is a topical absorbable hemostat consisting of a flowable gelatin matrix (paste). This kit also contains components necessary for preparing and applying the matrix during surgery. SURGIFLO Hemostatic Matrix, mixed with sterile saline or thrombin solution, is indicated in
| Date | Nov 16, 2012 (initiated Aug 21, 2012) |
| Status | Terminated |
| Reason | SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging. |
| Root cause | Packaging |
| Action | Ethicon Inc. sent an Urgent Product Information letter dated August 21, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their inventory to determine if they have any of the affected product. Affected product should be returned for replacement to Stericycle using the enclosed pre-paid shipping label. Customers with questions should call 1-877-384-4266 or contact their Ethicon Biosurgery Sales Representative. For questions regarding this recall call 877-384-4266. |
| Quantity | 45,200 |
| Distribution | Worldwide Distribution - USA (nationwide) and the countries of Russia, China, and the European Union. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.