Recall Z-0133-2013 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

DePuy Mitek, Inc., a Johnson & Johnson Co.

DePuy Mitek VAPR Hook Electrode 3.5mm Hook Electrode with Integrated Handpiece Catalog Number: 227305 The DePuy Mitek VAPR Electrodes for use with the VAPR Radiofrequency System are intended for resection, ablation, excision, hemostasis of blood vessels, and coagulation of soft tissue in patients requiring arthroscopic surgery of the knee, shoulder, ankle, elbow, and/or wrist.

DateOct 24, 2012 (initiated Mar 12, 2012)
StatusTerminated
ReasonProduct may have wrong large product identification label with symbols on the box indicating a S90 VAPR Electrode instead of the VAPR 3.5 hook electrode
Root causeError in labeling
ActionThe firm, Depuy Mitek, sent a "Medical Device Labeling Alert" letter dated March 14, 2012 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect your inventory to determine if any inventory of product code 227305 lot with the number M1108086 that is incorrectly labeled; review the following label information and the attached examples of correct and incorrect labels; and complete and return the Business Reply Form via fax to: 1-508-828-3750. To return the product call 1-877-379-4871 for an RGA and replacement product and
Quantity376 units
DistributionNationwide distribution: USA including states of: AL, CA, CT, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, NJ, NY, OH, OR, TN, TX, UT, VA, and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0826433.5 MM SIDE EFFECT ELECTRODE WITH INTEGRATED HANDPIECE, MODEL 227301, 3.5 MM ANGLED SIDE EFFECT ELECTRODE WITH INTEGRATE
Depuy Mitek, A Johnson & Johnson Company, Dec 19, 2008