ClearanceBrief
Recall Z-0395-2014 ยท NKB Thoracolumbosacral Pedicle Screw System

Synthes USA HQ, Inc.

Synthes Titanium Minimally Invasive Reduction Screw System Polyaxial Screw Body, Part Number 04.631.500 Non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation.

DateNov 26, 2013 (initiated Sep 7, 2012)
StatusTerminated
ReasonThis action is being initiated following a detailed investigation in response to reported complaints of post-operative locking cap loosening. Synthes has concluded that if the construct is mis-aligned during final tightening, the potential exists for locking caps to loosen post-operatively in certain situations.
Root causeUse error
ActionSynthes sent an Urgent: Medical Device Product Removal letter on September 7, 2012, to all affected consignees. Consignees were requested to immediately cease using the product, examine their inventory and remove the parts from inventory. If they had the parts, they were further instructed to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number, complete the verification section at the end of the letter and return the verification form. If they did not have the parts, they were requested to complete the verification section at the end of the letter and return the verifica
Quantity757
DistributionNationwide Distribution including CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113044SYNTHES MIRS
Synthes Spine, Dec 21, 2011