Recall Z-0112-2013 ยท JSS Kit, Identification, Enterobacteriaceae
Advandx Inc
AdvanDx GNR Traffic Light PNA FISH, In Vitro Diagnostic for Culture Identification Kit for Escherichia coli,and/or Klebsiella pneumoniae and/or Pseudomonas aeruginosa Catalog number: KT011. GNR Traffic Light PNA FISH is a multicolor, qualitative nucleic acid hybridization assay intended for the identification of Escherichia coli,and/or Klebsiella pneumoniae and/or Pseudomonas aeruginosa on smear
| Date | Oct 22, 2012 (initiated Sep 14, 2012) |
| Status | Terminated |
| Reason | GNR Traffic Light PNA Fish may have false positive results with Pseudomonas putida and Pseudomonas fulva |
| Root cause | Device Design |
| Action | AdvanDx sent a "GNR Traffic Light PNA FISH (KT011) Package Insert Replacement" letter dated September 14, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. All customers received an Acknowledgement Form to complete and return to AdvanDx so that the effectiveness of the replacement can be evaluated. Contact AdvanDx Technical Support at 1-866-376-0009 for questions regarding this notice. |
| Quantity | 187 kits |
| Distribution | Worldwide Distribution-USA (nationwide) and the countries of Austria, Denmark, Italy, Germany, Finland, and UK. |