Recall Z-1387-2013 ยท OAR Injector, Vertebroplasty (Does Not Contain Cement)
Orthovita, Inc., dBA Stryker Orthobiologics.
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System, 4", Part Number 2090-0502 Product Usage: Facilitate placement of bone cement in weakened or diminished bone
| Date | May 24, 2013 (initiated Nov 10, 2012) |
| Status | Terminated |
| Reason | There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use. |
| Root cause | Labeling design |
| Action | Stryker sent an Urgent Medical Device Correction letter dated November 10, 2012 to all affected customers. The letter identified the affected product, problem issue, potential hazards and actions to be taken. Customers were instructed to immediately review this notification and the attached IFU. Customers were asked to complete and sign the enclosed Business Reply Form and fax a copy to 1-484-323-8803. For questions contact Stryker Orthobiologics. |
| Quantity | 78 |
| Distribution | USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.