Recall Z-2464-2012 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

Siemens Magnetom Skyra, Verio, Spectra and Biograph MMR Nuclear Magnetic Resonance Imaging System and Tomographic Imager

DateSep 26, 2012 (initiated Sep 12, 2012)
StatusTerminated
ReasonWhen positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the head-foot direction, the coil can heat up in the area of the electronic housing (white plastic lids).
Root causeDevice Design
ActionSiemens Healthcare sent a Customer Safety Advisory Notice dated September 5, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter included updated instruction to all customers. For questions regarding this recall call 610-219-6300.
Quantity370
DistributionNationwide Distribution - USA including Washington, DC and AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072237MAGNETOM VERIO
Siemens Medical Solutions USA, Inc., Oct 10, 2007
K101347MAGNETOM AERA; MAGNETOM SKYRA
Siemens Medical Solutions USA, Inc., Oct 1, 2010
K103429BIOGRAPH MMR
Siemens Medical Solutions USA, Inc., Jun 8, 2011
K121160MAGNETOM SPECTRA
Siemens Medical Solutions USA, Inc., Jul 16, 2012