Recall Z-0006-2013 ยท FGO Dislodger, Stone, Flexible

US Endoscopy Group Inc

Endoscopic retrieval device. The disposable Raptor grasping device in intended to retrieve foreign bodies or resected tissue specimens in endoscopic procedures of the upper and lower gastrointestinal tract or biliary tract.

DateOct 3, 2012 (initiated Nov 14, 2011)
StatusTerminated
ReasonWhen the outer catheter is either coiled or in a contorted configuration outside of the endoscope and the patient, and excessive force is applied to the handle, the jaws may become inoperable and may become stuck in the open condition.
Root causeDevice Design
ActionUS Endoscopy initiated a Voluntary Product Recall letter initiated November 11, 2011. The letter identified the product the problem and the action needed to be taken by the customer. US Endoscopy will provide a free replacement or grant your facility a credit toward a future purchase for every device that is returned to US Endoscopy or verified as having been used or destroyed. For further questions please call 440-639-4494 ext. 378.
Quantity1153
DistributionWorldwide Distribution -- USA (nationwide) including the states of AL, AR, AZ, CA, FL, GA, HI, IA, IL, KY, LA, MD, ME, MI, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, and WI; and to the following countries: Argentina, Australia, Austria, Bahrain, Belgium, Brazil,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K962676DISPOSABLE ENDOSCOPIC RETRIEVAL DEVICE
United States Endoscopy Group, Inc., Oct 4, 1996