ClearanceBrief
Recall Z-0018-2013 ยท BSZ Gas-Machine, Anesthesia

Accutron Inc

Ultra PC% Cabinet Mount Package G, Model #36600, Serial # 13593, 13594, 13595, 13596, 13597, 13598, 13599, 13600, 13602, 13605, 13610, 13611, 13612, 13615, 13616, 13617, 13621, 13622, 13624, 13625, 13627, 13628, 13633, 13634, 13635, 13636, 13637, 13638, 13639, 13640, 13642, 13643, 13644, 13647, 13650, 13669, 13675, 13673, 13678, 13679, 13680, 13683, 13685, 13687, 13691, 13693, 13694, 13695, 13696

DateOct 9, 2012 (initiated Apr 29, 2012)
StatusTerminated
ReasonAccutron is recalling the Flowmeter (Analgesia Gas Machine) because when it flows Nitrous Gas without Oxygen gas being turned on, it functions in an incorrect manner.
Root causeDevice Design
ActionAccutron sent an Urgent Advisory Notice letter dated April 6, 2012 to all their customers who purchased the Ultra PC% Cabinet Mount Flowmeter. The letter informed the customers of the problem identified and actions to be taken. Customers were instructed to discontinue use or to continue use of the device only with extreme care until a replacement can be installed. Extreme care includes the use of the titration method for nitrous oxide administration coupled with close monitoring of both the patient and the actual gas flow through the flowtubes. For question contact Accutron at (800) 531-2
Quantity220 units total
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052335ANALGESIA GAS MACHINE, MODEL DIGITAL ULTRA
Accutron, Inc., Nov 23, 2005