Recall Z-2494-2012 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Philips Healthcare Inc.

Philips Digital Diagnost with patient carriage Model: 712050 with serial number range from 00 000 00 to 09 000 543. Product Usage: Stationary radiographic System

DateSep 28, 2012 (initiated Sep 12, 2012)
StatusTerminated
ReasonHook does not securely hold the footplate in vertical position, causing the footplate to disengage and may fall on patient or operator
Root causeDevice Design
ActionPhilips Healthcare sent an Urgent Field Safety Notice letter dated August 15, 2012 to all affected customers. The letter identified the affected product, problem and the actions to be taken to avoid or minimize the occurrence of the issue. Customers were advised to follow the instruction for Use and that Philips plans to replace the old hook with one that has a new design. For questions contact your local Philips representative 1-800-722-9377.
Quantity536 units
DistributionWorldwide Distribution - US Nationwide and the countries of :Australia, Austria, Bahrain, Belgium, Canada, Chile, China, Czech Republic, Denmark, Finland, , France, Germany, Ireland, Italy, Japan, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Russia, Saudi Arabia, Sweden, Slovakia, South
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982795PHILIPS BUCKY VISION
Philips Medical Systems, Inc., Nov 24, 1998