Recall Z-0037-2013 ยท KWQ Appliance, Fixation, Spinal Intervertebral Body

Medtronic Sofamor Danek USA Inc

ATLANTIS(R) Anterior Cervical Plate System FIXED ANGLE SCREW MODULE, REF 176-504, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used.

DateOct 10, 2012 (initiated Aug 30, 2012)
StatusTerminated
ReasonLines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.
Root causeNonconforming Material/Component
ActionMedtronic conducted a conference call on August 30, 2012 and sent an "URGENT VOLUNTARY RECALL" notice dated September 3, 2012 to all affected customers. The notice identified the product, problem, and actions to be taken by the customers. A Recall Questionnaire Form was attached to the letter for customers to complete and return. Contact your Medtronic Sales representative or the firm's Global Quality Department at 800-876-3133, ext. 3197 for questions regarding this notice.
Quantity1014 units
DistributionWorldwide Distribution-USA (nationwide) and the countries of India, Netherlands, Hong Kong, S. L Korea, Singapore, Canada, Australia, Japan, and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021461ATLANTIS VISION ANTERIOR CERVICAL PLATE SYSTEM
Medtronic Sofamor Danek, Jul 22, 2002
K063100ATLANTIS ANTERIOR CERVICAL PLATE SYSTEM, TRANSLATIONAL PLATE
Medtronic Sofamor Danek, Inc., Feb 23, 2007
K081038ATLANTIS ANTERIOR CERVICAL PLATE SYSTEM
Medtronic Sofamor Danek, Aug 15, 2008