Recall Z-0412-2013 ยท KWP Appliance, Fixation, Spinal Interlaminal

SpineFrontier, Inc.

Spine Frontier Inspan compressors Part Number: 11-60004. Inspan compressors provide pressure to drive spikes of Inspan implants into the spinous process bone, resulting in fixation of the implant.

DateNov 21, 2012 (initiated Aug 28, 2012)
StatusTerminated
ReasonInspan Compressors may break at weld causing a loss of compression
Root causeDevice Design
ActionSpine Frontier issued e-mail notices to customers on May 14, 2012 advising of the potential for breakage. Units were replaced with weld free design.
Quantity17 units
DistributionWorldwide Distribution-USA (nationwide) including PR and the states of FL, MA, MO, OK, and TX, and the country of Mexico.