Recall Z-0186-2013 ยท LLZ System, Image Processing, Radiological

Vital Images, Inc.

Software for Vitrea, Vitrea fX, and Vitrea Enterprise Suite (VES): VPMC-09293C Vitrea . Model Number Vitrea 5.2, 6.0, 6.1, and 6.2 VPMC-09346B Vitrea fX Model Number Vitrea fX 3.1, 5.2, 6.0, 6.1, and 6.2 VPMC-11731A Vitrea Enterprise Suite Model Number Vitrea Enterprise Suite 1.3, 6.0, 6.1, and 6.2. The software is a medical diagnostic system that allows the processing, review, analysis

DateNov 1, 2012 (initiated Aug 13, 2012)
StatusTerminated
ReasonVital Images has found two potential errors in calcium score values in restored snapshots of studies generated through use of Vitrea or Vitrea Enterprise Suite. This issue affects Version 3.1 of Vitrea fX; Version 5.2 of Vitrea; Versions 6.0, 6.1 and 6.2 of Vitrea and Vitrea fX; and Versions 1.3, 6.0, 6.1 and 6.2 of Vitrea Enterprise Suite; and all updates to those versions.
Root causeSoftware design
ActionVital Images sent an "URGENT DEVICE CORRECTION" letter dated August 13, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter gave advise on how to resolve the two issues and advised consignees to immediately inform all potential users. Contact Customer Support at 1-800-208-3005 for questions regarding the notice.
Quantity2088 (1133 USA)
DistributionWorldwide Distribution-USA (nationwide) including DC and Puerto Rico, and the countries of ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, BELARUS, BELGIUM, BRAZIL, BULGARIA, CANADA CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, FINLAND, FRANCE, GREAT BRITAIN, GEORGIA, GERMANY, GHA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032748VITREA2, VERSION 3.4 MEDICAL IMAGE PROCESSING SOFTWARE
Vital Images, Inc., Oct 29, 2003
K052632VITREA2 VERSION 3.8 MEDICAL IMAGE PROCESSING SOFTWARE
Vital Images, Inc., Dec 15, 2005