Recall Z-2486-2012 ยท HXX Screwdriver
Biomet, Inc.
Optilock T15 AO Driver (3.5mm). Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, r
| Date | Sep 28, 2012 (initiated Aug 23, 2012) |
| Status | Terminated |
| Reason | Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw |
| Root cause | Mixed-up of materials/components |
| Action | Biomet sent an "URGENT MEDICAL DEVICE RECALL NOTICE'' dated August 23, 2012 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. A Fax Back Response Form was included for customers to complete and return via fax to 574-372-1683. Contact the firm at 574-372-1570 for questions related to this recall. |
| Quantity | 363 units |
| Distribution | Worldwide Distribution-USA (nationwide) including the states of AL, GA, MT, TA, AZ, WA, CA, TA, IN, NY, NJ, VA, MA, MI, MD, PA, KY, NC, UT, NH, FL, OK, CO, LA, OH, MO and SD, and the countries of Japan, Netherlands, Australia, and Costa Rica. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.