Recall Z-0718-2014 ยท HTW Bit, Drill

The Anspach Effort, Inc.

Anspach Single Use Sterile Bone Cutting Burs and Irrigation Clips. Rx Only

DateJan 11, 2014 (initiated Sep 9, 2011)
StatusTerminated
ReasonRecall was initiated due to the observation of separation between clear layers of the pouch during real time aging studies. As a result, it is possible to have a breach in the sterile barrier.
Root causePackaging change control
ActionRecalling firm notified direct accounts by letter on 09/09/11. Direct accounts were instructed to screen their inventory immediately for product packaged with the Tyvek Pouch/Tube Configuration, complete the reply form, and return the products to the firm.
Quantity1,805,022 cutting burs and 55,026 irrigation clips
DistributionWorldwide Distribution - USA (nationwide) and the countries of Japan, Canada, Nigeria, Guam, China, Czech Republic, Thialand, Ecuador, Lithuania, Argentina, Turkey, Cayman Island, Jamaica, India, Jordan, Germany, Brazil, Mexico, Israel, Cameroon, Switzerland, South Africa, Greece, Finland, Taiwan, H
FDA recordOpen on fda.gov

510(k) clearances named in this recall