ClearanceBrief
Recall Z-2484-2012 ยท HWT Template

Zimmer, Inc.

00780304001 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780304002 MIS 12/14 Slotted Femoral Head Provisional +0X36 00780304003 MIS 12/14 Slotted Femoral Head Provisional +3.5X36 00780304004 MIS 12/14 Slotted Femoral Head Provisional +7X36 00780304005 MIS 12/14 Slotted Femoral Head Provisional +10.5X36 The provisional heads replicate final femoral implant geometry and are used to trial ra

DateSep 28, 2012 (initiated Apr 1, 2011)
StatusTerminated
ReasonComplaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
Root causeDevice Design
ActionZimmer, Inc. sent an E-mail letter dated April 21, 2011, to all affected customers in the US. A June 2011 E-mail was sent to all customers outside the US. The letters requested Inventory & Operations Managers to check product for proper fit /function, complete the return form with worn provisional heads to Zimmer Inc, Attention Product Service, 1777 West Center St. Warsaw, IN 46580. For questions regarding this recall call 574-372-4753.
Quantity231,641 all lots, all sizes
DistributionWorldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.