Recall Z-0009-2013 ยท OMP Negative Pressure Wound Therapy Powered Suction Pump

Accuro Medical Product LLC

Wound Pro Apex Negative Pressure Wound Therapy Powered Suction Pump; Accuro Medical Products LLC, 2854 Hitchcock Ave, Downers Grove, IL 60515; part number 84635400182, sold inside the Apex Negative Wound Therapy system, part number 84635400181 Promotes wound healing through drainage and removal of infectious material and other fluids from the wound site using continuous and/or intermittent negati

DateOct 4, 2012 (initiated Aug 31, 2012)
StatusTerminated
ReasonAccuro has received reports of some Apex Pumps either failing to operate or malfunctioning, at times without notifying the user of such failure or malfunction on the unit's display screen.
Root causeDevice Design
ActionThe firm, Accuro, sent an "Urgent: Medical Device Recall" letter dated August 30, 2012 to the customers via e-mail on August 31, 2012. The letter describes the product, problem and actions to be taken. The customers were instructed to remove all devices from use immediately; to immediately complete and return the enclosed Accuro response form via email to; (customer.service@accuromedical.com) or fax to: (1-800-866-0545) . Note: the form still needs to be returned even if you no longer have or use the device. Upon receipt, Customer Service will call the account with instructions to arrange for
Quantity919 units
DistributionWorldwide distribution: USA (nationwide) and country of: Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100823WOUND PRO APEX
Accuro Medical Products, LLC, Jan 24, 2011